CHLO-206A Digital Chlorophyll Sensor
The Apure CHLO-206A is an in-water fluorescence sensor for monitoring chlorophyll-related changes. Interpret its chlorophyll-a estimate with the supplied calibration and site reference samples, and plan routine optical cleaning and verification.
- Measurement Precision
- RS485
- Automation
The CHLO-206A excites algal pigments and measures the emitted fluorescence. A configured relationship reports a chlorophyll-a estimate; the fluorescence signal is not itself a direct count of algal cells or a laboratory extraction result.
The listed optical compensation and ambient-light suppression functions concern instrument response. Confirm the supplied implementation; these functions do not remove every effect of species composition, particle interference, temperature or light-induced fluorescence quenching.
Calibrate and interpret the fluorescence result
- Use the reference medium, zero/span procedure and acceptance criteria supplied for the CHLO-206A. A dye check can assess response or drift, but it does not alone establish the relationship to every natural algal community.
- Collect representative water samples at the sensor location for an appropriate laboratory chlorophyll-a analysis. Use paired results to check the concentration relationship across the intended conditions.
- Record time, depth, light conditions and temperature. Strong-light physiological quenching or a change in species can alter fluorescence without a matching change in pigment concentration.
- Inspect and clean the optical surface as directed. Verify after cleaning or service and set a site maintenance schedule; online operation does not mean maintenance-free operation.
The listed 0–400 µg/L range and other values remain configuration references. Confirm the concentration basis and accuracy conditions for the ordered sensor. The IP68/depth description requires the supplied installation and connector conditions; a temperature sensor is not universal compensation for all optical effects.
Technical parameter
- 0-400ug/L
- Resolution: 0.1ug/L
- 0-50 °C; IP68 protection, water depth within 20 meters
- Built-in temperature sensor
- Support RS-485, Modbus/RTU protocol
- Convenient, fast, stable and easy to maintain
| Brand | Apure |
|---|---|
| Model | CHLO-206A |
| Measuring range | 0~400ug/L |
| Resolution | 0.1 ug/L,0.1℃ |
| Accuracy | ±3%,±0.5℃ |
| Digital output | Isolated protection RS485 signal output |
| Cable length | 5 meters, other lengths can be customized |
| Temperature | 0~50℃ |