NO3-201 Digital Nitrate Nitrogen Sensor

The Apure NO3-201 uses a nitrate ion-selective electrode for online water monitoring. Confirm the selected calibration and whether mg/L is reported as nitrate ion (NO3−) or nitrate as nitrogen (NO3-N).

  • Measurement Precision
  • High stability
  • Long life

The listed NO3-201 uses a PVC-membrane nitrate ion-selective electrode with a temperature sensor. Its electrical response depends on nitrate-ion activity; calibration, the reporting basis and the sample matrix determine how that response is expressed as a concentration.

Set up a comparable nitrate result

  • Specify NO3− or NO3-N explicitly on the calibration record, controller and digital register. These are different mass-reporting bases and cannot be compared as identical mg/L readings.
  • Use the supplied electrode-conditioning and calibration procedure with suitable nitrate standards. After calibration, compare against an independent sample reference on the same reporting basis.
  • Check interfering anions, ionic strength, electrode condition and water chemistry. The impact of chloride or other ions depends on the actual electrode and concentrations; obtain the supplied selectivity information.
  • Temperature compensation does not establish selectivity or replace matrix validation. Recheck when the process composition or expected concentration changes.

Install and integrate

  • Confirm the listed 3/4-inch NPT mounting, wetted materials, cable and approved immersion or flow arrangement. Keep the sensing membrane representative of the water and accessible for maintenance.
  • The existing temperature row also lists a pressure value. Verify the actual temperature and pressure rating for the supplied probe, connection and process before installation.
  • For RS485 Modbus RTU, obtain the actual power and wiring requirements, register map, unit basis and serial settings. A nitrogen-reporting calculation must be verified rather than inferred from the page title.

The original Resolution entry, ±5%F.S., is a percentage-of-full-scale expression and does not define a concentration display increment. It is retained below as an unverified entry; confirm actual resolution, the basis of the listed 0~1000mg/L range and the accuracy conditions for the ordered version.

BrandApure
ModelNO3-201
Measuring range0~1000mg/L
Resolution entry — unverified±5%F.S.
Accuracy±5%F.S.
Digital outputIsolated protection RS485 signal output
Cable length10 meters, other lengths can be customized
Temperature0~50℃;<0.3MPa