NHN Online Ammonia Nitrogen Sensor

The Apure NHN sensor, listed as NHN-206, uses an ammonium ion-selective electrode for online water monitoring. Specify whether the required result is ammonium, ammonium as nitrogen, or total ammonia nitrogen before configuring the system.

  • Measurement Precision
  • High stability
  • Long life

The listed PVC-membrane electrode responds to ammonium-ion activity through a membrane potential. Calibration relates that response to the selected reporting scale. Temperature compensation does not by itself correct ionic-strength effects, interfering ions or the chemical distribution between NH4+ and NH3.

Required resultWhat must be established
Ammonium ion, NH4+Use standards and output units defined for ammonium ion.
Ammonium as nitrogen, NH4-NConfirm the nitrogen mass basis in the calibration, display and digital register.
Total ammonia nitrogen, NH3-N plus NH4-NAn ammonium electrode alone does not directly measure both forms. Verify the required pH/temperature inputs and validated calculation or use an appropriate reference method.

Calibration and installation

  • Select the supplied range, standards and conditioning procedure for NHN-206. Verify a separate reference sample after calibration, with sample timing and reporting basis matched.
  • Assess interfering ions, particularly potassium for an ammonium ISE, and changes in water chemistry. Obtain the model-specific selectivity and matrix-adjustment instructions; do not assume universal compensation is installed.
  • The listed 100 KPa (1 Bar) reference-system statement does not establish the permitted pipe pressure. Confirm process-pressure, immersion, materials and sealing limits for the complete probe.
  • The listed mounting is 3/4 NPT with IP68 protection. Confirm the actual thread, cable/connector and installation conditions; provide access for electrode cleaning, reference checks and replacement.
  • The listed interface is RS485 Modbus RTU. Match supply, wiring, register units and serial settings to the receiving device.

The entry previously labelled Resolution is ±0.5℃. Its unit does not specify concentration resolution; the actual concentration display increment must be confirmed for the supplied NHN-206. The original value is retained below as an unverified entry, not reassigned to a different specification. Confirm the selected range and accuracy conditions separately.

BrandApure
ModelNHN-206
Measuring range0~10.00mg/L(0~100.00mg/L)
Resolution entry — unverified±0.5℃
Accuracy±10% or ±1mg/L
Digital outputIsolated protection RS485 signal output
Cable length5 meters, other lengths can be customized
Temperature0~40℃