BOD-201X BOD Sensor

The Apure BOD-201X provides an online optical estimate for tracking BOD-related changes. Validate its relationship to the required laboratory BOD method for the water being monitored.

  • Measurement precision
  • High stability
  • Low power design

Laboratory BOD5 measures oxygen consumption during a defined biological incubation, commonly five days at 20°C. An immediate optical reading does not carry out that incubation. It can support trend monitoring when an appropriate relationship to the laboratory result has been established.

Optical estimate and laboratory reference

Confirm the supplied BOD-201X optical method, measured channel and conversion algorithm. UV absorbance and fluorescence are different optical methods; neither becomes a direct biological-incubation measurement simply because the output is labelled BOD. Changes in the mixture of biodegradable and other compounds can change the relationship.

Listed indicationInterpretation / check
mg/L equiv. KHPA KHP-equivalent calibration expression does not by itself establish a BOD5 result or method equivalence. Confirm how the supplied instrument defines this channel.
NTUAn optical turbidity channel uses a different measurand and unit from BOD. Confirm whether this is an available channel in the supplied version.
Online BOD estimateCompare paired site samples with the chosen BOD5 or other stated reference method. Validate the relationship and its applicable process range.

Commission and use

  • Agree the reference method, sample handling and time/location of paired samples. Keep calibration and independent verification samples distinct, and record the relationship and error over the intended range.
  • Reassess after a source or process change. Review bubbles, particle loading, optical-window fouling and changes in organic composition before accepting a shift in the estimate.
  • Confirm installation, supply, actual temperature/pressure limits and the cleaning arrangement. Inspect, clean and verify according to the supplied procedure.
  • The listed RS485 Modbus RTU output needs a compatible receiver and confirmed register units. Use online data for the validated monitoring purpose; retain the required laboratory method for any reporting that calls for it.

The original KHP-equivalent and NTU ranges, resolution and accuracy figures are retained as listed references. Confirm the actual channel configuration, calibration basis and performance definition for the supplied BOD-201X. These entries are not proof of BOD5 equivalence or a guaranteed site-correlation accuracy.

BrandApure
ModelBOD-201X
Measuring range0-100.0mg/L equiv. KHP, 0-100.0NTU
Resolution0.1NTU, 0.1mg/L
Accuracy±5% FS
Digital outputRS485 Modbus RTU
Cable length10m, other can be customized
Temperature0℃ to 50℃