COD-208 UV Cod Sensor

The Apure COD-208 uses UV optical measurements to estimate COD-related changes in water. Select the listed range and validate the optical-to-COD relationship against representative samples from the intended process.

  • Measurement Precision
  • High stability
  • Low power design

The listed COD-208 design uses a 254 nm UV channel and a reference channel for compensation. UV absorbance responds to UV-absorbing constituents; it does not directly measure oxygen consumed in a chemical digestion. A COD estimate therefore depends on the water matrix and the configured relationship to a reference COD method.

Learn how laboratory COD, UV254/SAC254 and online optical COD estimation differ. Document the selected reference method and site-validation plan before using the reported optical COD estimate for an alarm or control decision.

Calibrate for the water being monitored

  • Specify the water source, expected range and reference laboratory method, including whether samples are filtered or unfiltered. Optical and laboratory results should represent the same sample fraction.
  • Collect paired optical readings and laboratory COD results across the relevant process conditions. Validate the relationship on additional samples, record residual differences and reassess it when the wastewater composition changes.
  • Check optical windows, bubbles, suspended particles and deposits before changing calibration. A reference-channel correction can reduce certain effects; it does not remove every interference or compensate for all changes in organic composition.

Installation and use

  • Use for trend monitoring and process assessment in compatible surface-water or wastewater installations after site validation. A reading alone does not establish discharge compliance, drinking-water safety or total pollutant content.
  • Keep the optical path immersed and representative of the process. Confirm mounting, flow, materials and the pressure/temperature limits of the supplied assembly.
  • The listed cleaning brush assists with fouling control. Inspect and clean the windows, check brush condition and verify readings with the supplied procedure; automatic cleaning does not make the sensor maintenance-free.
  • For the listed RS485 output, obtain the actual application protocol, register map, engineering units and receiving-device settings. RS485 is a digital interface, not an analog current output.

The two listed COD ranges are alternatives to confirm for the supplied version. The accuracy entry “±5%F.S. / 0.1mg/L” is retained as listed; confirm the error definition, calibration basis and applicable sample conditions before specifying performance.

BrandApure
ModelCOD-208
Measuring range0~200mg/L or 0~500mg/L
Resolution0.1mg/L
Accuracy±5%F.S. / 0.1mg/L
Digital outputIsolated protection RS485 signal output
Cable length5 meters, other lengths can be customized
Temperature0~45℃