TN-300 Total Nitrogen Analyzer
The Apure TN-300 is an online analyzer for total-nitrogen monitoring. Confirm the supplied analytical method, sample fraction and calibration before interpreting the result as total nitrogen in the intended water body.
- Stable performance
- Quick installation
- Long life
Total nitrogen includes the relevant inorganic and organic nitrogen forms in the sample, expressed on a nitrogen mass basis. A nitrate-only or ammonium-only measurement is not a complete total-nitrogen determination. The sampling and conversion procedure must match the fraction being reported.
| Required result | What must be established |
|---|---|
| Dissolved nitrogen | The filtered-water fraction and the method for the relevant dissolved nitrogen forms must be specified. |
| Total nitrogen in whole water | The method must account for dissolved and particulate nitrogen. Suspended material and incomplete conversion can bias an apparent total result. |
| Comparison with nitrate, nitrite, ammonia or TKN | Match the analytical definitions and units; a single species or TKN alone is not interchangeable with total nitrogen. |
Confirm the method and data quality
- Obtain the current TN-300 method description: sample collection, any conversion or digestion, optical detection, reagents and calibration. The previous colour-reaction wording does not by itself identify a validated method or its recovery.
- Where the supplied method converts nitrogen forms to a common measurable species, verify conversion as well as the final detector response. Use blanks, independent checks and paired representative samples with the chosen laboratory reference.
- Assess particles, organic matter, salinity, carry-over and expected concentration. A method suitable for clear water does not automatically provide unbiased total-nitrogen results in every sediment-rich or coastal sample.
- Document the measurement cycle, sample timing, status flags, invalid readings and maintenance events. Nutrient trends support assessment; a TN value alone does not establish the timing or toxicity of an algal bloom.
Power, deployment and listed references
The existing description lists 24 VDC supply, 0.001 mg/L resolution and IP68 protection. These references are retained; confirm the current supply/load profile, reporting increment and complete assembled ingress rating. They do not establish battery autonomy, accuracy at the display increment, or that every installation can be continuously submerged.
For the listed RS485 Modbus RTU connection, obtain the supplied wiring, register map, nitrogen units and timing/status information. Buoy, shore and ship deployment also require the appropriate sampling arrangement, power budget, reagents and waste handling where applicable, and access for servicing.
Original range, accuracy, external temperature/pressure and the additional references in the description are retained. Confirm the actual analytical method, fraction, resolution, measurement cycle, conversion recovery, power, consumables and applicable performance conditions for the supplied TN-300; no unverified chemistry or specification has been substituted.
| Brand | Apure |
|---|---|
| Model | TN-300 |
| Measuring range | 0~5mg/L |
| Accuracy | 0.05mg/L or ±10% |
| Digital output | RS485 Modbus RTU |
| Listed operating conditions | 0~50°C, 0~1Bar |