ACT Insertion Type Magnetic Flowmeter

The Apure ACT is an insertion electromagnetic flowmeter for a compatible conductive liquid in a full pipe. The probe senses velocity at its installed position. Accurate volumetric flow depends on the internal pipe diameter and the insertion and flow-profile settings specified for the supplied assembly.

  • Multi-functional Intelligent Converter
  • Rapid response and high stability
  • High-definition LCD Backlight Display

The ACT is an insertion-type electromagnetic flowmeter for a suitable conductive liquid application. An inserted sensing head detects the electrical response associated with liquid moving through a magnetic field. The controller relates that response to the configured flow measurement for the installation.

Installation determines measurement performance

  • Confirm liquid conductivity, pipe internal diameter, expected flow interval and the model’s installation requirements. Pipe geometry and velocity profile affect an insertion measurement.
  • Use the specified fitting, insertion depth, orientation and upstream/downstream straight-run conditions. Keep the pipe full and avoid gas pockets or deposits at the measuring head.
  • The inserted sensing head and electrodes contact the process liquid. Confirm compatibility of every wetted material and seal; insertion construction does not remove corrosion considerations.
  • Use the approved process isolation and pressure-release procedure for installation or removal. Installation without shutdown is permissible only for an explicitly suitable assembly and procedure, not for every insertion meter.
  • Match grounding, shielding, power, signal type and output scaling to the supplied ACT documentation. Do not copy another product family’s pH wiring or range settings.

Commission and verify

Record the model and serial number, pipe dimensions, installation geometry, configured units/range and outputs. Check the full-pipe condition and diagnostics, then compare with an appropriate reference or site verification procedure across the intended operating interval. Follow the ACT manual for any zero or calibration operation.

Specify the insertion version when requesting a quotation. Full-bore lined-meter size tables and performance claims cannot automatically be applied to an insertion assembly. Request the actual range, accuracy basis, pressure/temperature ratings and mounting drawing for the supplied ACT version.

BrandApure
ModelACT insertion-type electromagnetic flowmeter
MeasurementFlow of a compatible conductive liquid
Flow specification requiredConfirm velocity/flow-rate range, unit, pipe internal diameter, configuration and accuracy basis for the supplied insertion version.
Installation requiredConfirm fitting, insertion depth, straight runs, orientation, grounding and full-pipe requirements from the mounting instructions.
Process and electrical ratingsConfirm wetted materials, seals, pressure, temperature, power supply, interface and outputs for the ordered assembly.