SR-510 Inline PTSA Sensor

The Apure SR-510 is an inline fluorescence sensor for PTSA tracer monitoring in compatible water-treatment systems. Use its PTSA reading with the documented tracer-to-treatment-product relationship when configuring dosage feedback.

For fluorescein, NDSA or another tracer, confirm a suitable calibrated optical channel and the ordered configuration. A PTSA reading alone does not identify or quantify every fluorescent tracer.

  • Measurement precision
  • High stability
  • Low power design

PTSA fluorescence is detected after the tracer absorbs excitation light and emits light at a different wavelength. The calibrated response indicates PTSA concentration within the validated range. It does not directly measure fluid flow, locate an oil reservoir or determine every active ingredient in a treatment chemical.

Required informationUse of the PTSA result
PTSA concentrationCheck the supplied calibration and compare with an independent PTSA reference.
Treatment-product doseUse the known PTSA content of that formulation and the correct units. A change in product or tracer ratio requires a new relationship.
Active inhibitor or treatment performanceTracer concentration does not independently prove the remaining active inhibitor, corrosion control or scale-control performance. Use the relevant checks.

Commission the monitoring loop

  • Choose a representative, well-mixed sampling point. Confirm flow-cell installation, materials, pressure, flow and temperature limits for the supplied SR-510.
  • Use the supplied zero/span procedure and appropriate PTSA standards. Verify after calibration, and review background fluorescence, coloured water, particles, bubbles and optical-window deposits if the result changes.
  • Confirm the controller, output scaling, alarm settings and response to loss of flow or invalid readings before enabling dosing. The sensor alone does not establish a complete dosing-control package.
  • Inspect and clean the optical surfaces using the supplied method. Set verification and maintenance intervals for the actual water and chemical programme.
  • The listed outputs include RS485 Modbus RTU and 4–20 mA. Confirm which are supplied, their wiring, digital register map and analog scaling; an analog current signal is distinct from the digital protocol.

Original range, response, accuracy, cable and temperature values are retained. Confirm the response-test conditions, pressure/flow limits and output version in the current SR-510 documentation. The list of other tracers does not establish that every channel is included.

BrandApure
ModelSR-510
Measuring range0-200ppb
Response<30s
Accuracy±1%FS
Signal outputsRS485 Modbus RTU 4-20mA
Cable length5 meters
Temperature4~50℃