TS-620 Turbidimeter Online Turbidity Meter

The Apure TS-620 is an online turbidity meter for use with a matched optical sensor in water and wastewater monitoring. Select the sensor option and working interval, then confirm the measurement method, reporting unit and required outputs.

  • Easy to operate
  • Widely used
  • Automation

The TS-620 monitors turbidity with its matched optical sensor. Turbidity is an optical measurement influenced by particle properties and the measuring arrangement; it is not a direct measurement of suspended-solids mass or drinking-water safety.

Install and check the optical measurement

  • Place the sensor in representative water within the approved temperature and pressure limits. Avoid trapped bubbles, deposits, direct aeration plumes and unapproved reflections from nearby walls or the vessel bottom.
  • Clean the optical window using the supplied procedure, check for scratches and fouling, and verify with the specified standards or reference samples after service.
  • Record the sensor/controller identities, measurement method, unit, calibration reference, temperature and installation. A change in colour, particle properties or sampling location can change the optical response.

The TD-400 and TD-5000 entries below are separately listed options. Confirm the selected sensor and interval; do not combine them into one universal range. The existing relay descriptions differ between entries, so obtain the actual terminal drawing and relay/output list for the ordered controller.

CHARACTERISTICS:

Operation and integration

  • Maintain the specified sample and installation conditions; changes in flow, pressure, bubbles or particle distribution can affect an optical reading.
  • Confirm the supplied display values and unit. A percentage or current-output display is not an independent measurement of suspended-solids mass.
  • For the listed 4-20mA, relay and RS485 options, use the actual terminal drawing and communication specification. Confirm enabled functions before control-system integration.
  • Follow the documented two-point calibration procedure using the specified references. Keep a separate verification result and maintenance record.
Electrode measurement rangeResolutionPrecision
TD-400: 1-100.00ntu0.01NTU±1%FS
TD-5000: 20-5000ntu0.1NTU±1%FS
Medium temperature: 0-50℃Environmental humidity: relative humidity is less than or equal to 85%
Size of instrumentInstallation
242mm*184mm*135mm(H*W*D)wall

Confirm the ordered sensor/controller pair and the actual relay/output schedule. The existing entries use different relay descriptions; the table has been retained as listed rather than choosing an unverified count.

BrandApure
ModelTS-620
Precision±1%FS
Medium temperature0-50℃
Electrode measurement rangeTD-400: 1-100.00ntu TD-5000: 20-5000ntu
ResolutionTD-400: 0.01NTU TD-5000: 0.1NTU
Control interface4-Relay(high-low point trigger)
Environmental humidityrelative humidity is less than or equal to 85%
Signal isolation output1 set of 4-20 ma; 2 sets of alarm relays
Dimensions242×184×135mm