TP-300 Total Phosphorus Analyzer

The Apure TP-300 is an online phosphorus analyzer with a listed digestion and spectrophotometric workflow. Define the phosphorus fraction, sample handling and reference method required for the supplied configuration.

  • Stable performance
  • Quick installation
  • Long life

Total phosphorus and direct orthophosphate are different measurements. In a digestion-plus-colorimetry method, relevant phosphorus forms are converted to a measurable form before the optical determination. A direct orthophosphate reading alone does not establish total phosphorus.

Required resultSampling and method requirement
OrthophosphateA direct reactive-phosphorus measurement is distinct from a total-phosphorus determination.
Total dissolved phosphorusA filtered sample requires the specified conversion and measurement procedure; the result concerns the dissolved fraction.
Total phosphorus in whole waterThe sampling and conversion method must account for the relevant dissolved and particulate forms. Removing particles before analysis can change the fraction reported.

Listed method and supplied configuration

The existing description lists persulfate digestion at 120–124°C, conversion to orthophosphate and a molybdate/antimony/ascorbic-acid colour reaction. Confirm the current TP-300 procedure, reagents, reaction conditions and calibration. A digestion temperature is distinct from the external water or housing operating temperature; the listed value is retained as a method reference, not a newly verified process rating.

Specify and verify the complete installation

  • Confirm sampling, any filtration or homogenisation, conversion recovery, blank correction and calibration basis. Express results explicitly as phosphorus and match the sample fraction to the laboratory reference.
  • Verify independent quality-control samples and representative paired field samples, including relevant particles and the salinity expected at the site. A valid orthophosphate standard check alone does not test conversion of all phosphorus forms.
  • For buoy, shore or ship deployment, confirm the complete supplied equipment, power budget, measurement cycle, reagents, cleaning and waste handling where applicable. Do not infer reagent-free operation or unlimited autonomy from the term in-situ.
  • Confirm the actual external temperature/pressure limits, materials and installation arrangement. Set servicing and verification intervals for the water conditions.
  • The listed RS485 Modbus RTU connection needs the supplied register map, units, timing and validity/status information. A phosphorus reading supports nutrient assessment but does not independently predict a specific algal event.

Original range, error and external temperature/pressure entries are retained. Confirm the actual phosphorus fraction, method, measurement cycle, consumables, accuracy definition and environmental ratings for the ordered TP-300. This page does not certify compliance with a particular laboratory standard.

BrandApure
ModelTP-300
Measuring range0~2mg/L
Accuracy0.05mg/L or ±10%
Digital outputRS485 Modbus RTU
Listed operating conditions0~50°C, 0~1Bar