TP-300 Total Phosphorus Analyzer
The Apure TP-300 is an online phosphorus analyzer with a listed digestion and spectrophotometric workflow. Define the phosphorus fraction, sample handling and reference method required for the supplied configuration.
- Stable performance
- Quick installation
- Long life
Total phosphorus and direct orthophosphate are different measurements. In a digestion-plus-colorimetry method, relevant phosphorus forms are converted to a measurable form before the optical determination. A direct orthophosphate reading alone does not establish total phosphorus.
| Required result | Sampling and method requirement |
|---|---|
| Orthophosphate | A direct reactive-phosphorus measurement is distinct from a total-phosphorus determination. |
| Total dissolved phosphorus | A filtered sample requires the specified conversion and measurement procedure; the result concerns the dissolved fraction. |
| Total phosphorus in whole water | The sampling and conversion method must account for the relevant dissolved and particulate forms. Removing particles before analysis can change the fraction reported. |
Listed method and supplied configuration
The existing description lists persulfate digestion at 120–124°C, conversion to orthophosphate and a molybdate/antimony/ascorbic-acid colour reaction. Confirm the current TP-300 procedure, reagents, reaction conditions and calibration. A digestion temperature is distinct from the external water or housing operating temperature; the listed value is retained as a method reference, not a newly verified process rating.
Specify and verify the complete installation
- Confirm sampling, any filtration or homogenisation, conversion recovery, blank correction and calibration basis. Express results explicitly as phosphorus and match the sample fraction to the laboratory reference.
- Verify independent quality-control samples and representative paired field samples, including relevant particles and the salinity expected at the site. A valid orthophosphate standard check alone does not test conversion of all phosphorus forms.
- For buoy, shore or ship deployment, confirm the complete supplied equipment, power budget, measurement cycle, reagents, cleaning and waste handling where applicable. Do not infer reagent-free operation or unlimited autonomy from the term in-situ.
- Confirm the actual external temperature/pressure limits, materials and installation arrangement. Set servicing and verification intervals for the water conditions.
- The listed RS485 Modbus RTU connection needs the supplied register map, units, timing and validity/status information. A phosphorus reading supports nutrient assessment but does not independently predict a specific algal event.
Original range, error and external temperature/pressure entries are retained. Confirm the actual phosphorus fraction, method, measurement cycle, consumables, accuracy definition and environmental ratings for the ordered TP-300. This page does not certify compliance with a particular laboratory standard.
| Brand | Apure |
|---|---|
| Model | TP-300 |
| Measuring range | 0~2mg/L |
| Accuracy | 0.05mg/L or ±10% |
| Digital output | RS485 Modbus RTU |
| Listed operating conditions | 0~50°C, 0~1Bar |