CU-406 Copper Ion Selective Electrode
The Apure CU-406 is an online copper ion-selective sensor for compatible aqueous samples. Define the required copper fraction and sample method before using its reading for process control or comparison with laboratory results.
- Stable performance
- Quick installation
- Long life
A cupric ion-selective electrode responds through a membrane potential related to free Cu2+ activity. Calibration under defined conditions relates that signal to the reporting scale. It does not inherently measure all dissolved, complexed or particulate copper, and it does not remove copper from the water.
Choose the copper measurement task
- Specify free cupric ion, dissolved copper or a method-defined total copper result. These are different analytical tasks; a total-copper laboratory result is not automatically comparable with a direct free-ion reading.
- Obtain the actual electrode selectivity and compatible sample conditions. Other ions, complexing agents, ionic strength, pH and deposits can affect the response; the sensor is not interference-free.
- Use the supplied conditioning, standards and calibration procedure, then check an independent sample reference. Any extraction, digestion or other preparation must follow the selected method and material compatibility.
- Water and compatible process streams are the direct application. Soil, food, pharmaceutical or biological samples require a validated preparation and analytical method; this probe is not a direct air/soil analyser or a standalone safety certification.
Connect and maintain the digital sensor
- The listed interfaces are RS485 Modbus RTU and 4–20 mA. Confirm the supplied version, power, wiring, register units and analog span. Do not connect a digital or current-output probe directly to an ordinary pH/mV input as if it were a bare electrode.
- The Pt1000 entry describes listed temperature compensation, not the allowable process-temperature range. Obtain the actual temperature, pressure, pH and materials limits for the supplied assembly.
- Install in representative water with access for cleaning and calibration checks. Use the specified cleaning, conditioning and storage method, and record drift and reference results.
Original ranges, error figures, output entries and Pt1000 reference are retained. The existing description identifies a PVC-film construction; confirm the actual membrane and supplied electrode version. The temperature row is labelled Temperature compensation to match its existing content; no operating-temperature limit is inferred. Confirm the reporting basis and complete installation ratings with the current CU-406 documentation.
| Brand | Apure |
|---|---|
| Model | CU-406 |
| Measuring range | 0-100.00mg/L, 0-1000.0mg/L |
| Accuracy | ±10% or ±1mg/L, ±0.5℃ |
| Signal outputs | RS485 Modbus RTU & 4-20mA |
| Temperature compensation | Automatic Temperature Compensation (Pt1000) |