KOG-206 Optical Dissolved Oxygen Sensor

The Apure KOG-206 is an optical dissolved oxygen sensor using an oxygen-sensitive luminescent layer. Select the sensor and matched transmitter for the water type, installation and required control-system interface.

  • Visualization
  • High anti-jamming capacity
  • Far transmission distance

KOG-206 setup and calibration

The KOG-206 uses optical luminescence sensing: oxygen changes the response of an oxygen-sensitive layer. This optical route does not use the membrane/electrolyte polarization procedure of a polarographic probe.

  • Inspect and clean the optical sensing surface using the supplied instructions. Check the cap identity, condition and replacement procedure; do not apply an electrochemical electrolyte-filling procedure to the optical cap.
  • Allow temperature and readings to stabilize. Identify which temperature measurement and compensation functions are provided by the sensor and by the matched controller.
  • For the listed air or sample calibration, use the controller’s specified medium, stability criteria and sequence. Check local pressure and required salinity entries; do not assume either is measured or corrected automatically.
  • After calibration or cap replacement, perform the specified independent verification and record sensor/controller identities, cap details, reference conditions, settings and the result. Mount the sensor in representative water with access for cleaning.

The exact reference medium, cap interval, pressure/salinity functions and acceptance limits depend on the supplied configuration. See the dissolved oxygen sensor guide for comparison with polarographic sensors.

Features:

  • Optical sensing uses an oxygen-sensitive luminescent layer. Inspect its surface and follow the supplied instructions for cleaning and sensing-cap replacement.
  • The optical measurement does not consume oxygen. Mount the probe in representative water and follow the installation requirements for bubbles, fouling and sample exchange.
  • Match the listed NTC temperature-compensation arrangement to the supplied sensor and receiving system. Confirm where pressure and salinity corrections are applied.
  • Verify the factory calibration during commissioning. Set verification and calibration intervals for the application, sensing-cap condition and supplied procedure; cleaning, cap replacement or a failed verification may require recalibration.
  • For digital integration, match the supplied power, interface, protocol and data format to a compatible receiving device. Select a matching transmitter when its documented analogue outputs, relays or display functions are required.

Cable information

The cable colour key below is a reference for the listed configuration. Confirm the wire functions and sensor supply for the exact ordered probe before connection. Use each device’s own instructions when selecting sensor power, transmitter power and output options.

RS-485 signal output

  1. Red wire- Power(12-24V)
  2. Black wire-Ground(GND)
  3. Blue wire-485A
  4. White wire-485B
  5. Bare wire-shield
Examples of dissolved oxygen probe construction
Examples of different dissolved oxygen probe constructions. Select the optical KOG-206 and its compatible receiving system using the supplied model documentation.

The specifications below describe the listed sensor/transmitter configuration. Confirm the supplied sensor, transmitter, power supply, output, materials and cable in the current datasheet and quotation before ordering.

Measurement rangeDO:0-20 mg/L,0-20 ppm;
Temperature:0-45℃
Measurement AccuracyDO:±3% of the measured value; Temperature:±0.5℃
Pressure range≤0.3Mpa
CalibrationAir or sample calibration according to the supplied controller procedure
Main materialBody:SUS316L + PVC(fresh water),Titanium alloy(marine);O-ring:fluorine rubber;Cable:PVC
Power supplyAC:100-240VAC (50/60HZ)
DC:9~36VDC
Output2-way 4-20mA
OutputSetting three-way relays,programmed response parameter and response value.
Communication protocolRS485(customize)
Storage temperature-15-65℃
Measuring temperature0-45℃
DimensionsSensor:Dia55mm*L 342mm
Transmitter:142*142*162mm(L*W*H)
WeightSensor:1.85KG;Transmitter:1.35KG
Protective rateSensor:IP68/NEMA6P;Transmitter:IP65/ NEMA4X
Cable lengthStandard:10m, the maximum may be extended to 100m