KOG-202S Polarographic Dissolved Oxygen Sensor
The Apure KOG-202S is a polarographic dissolved oxygen sensor for measuring oxygen dissolved in water. Select it together with a compatible controller and confirm the supplied configuration’s operating limits and service requirements.
- Easy to operate
- Widely used
- Automation
KOG-202S setup and calibration
Use the procedure for the supplied KOG-202S probe and its matched controller. Polarographic DO measurement uses an applied polarization voltage and an oxygen-permeable membrane; it is different from optical luminescence sensing.
- Inspect the membrane, seals and electrolyte as directed by the supplied manual. Complete the required polarization after connection or service, and allow temperature and readings to stabilize.
- Check where temperature compensation is performed and which pressure and salinity inputs the controller requires. Temperature compensation alone does not establish automatic pressure or salinity correction.
- Use the specified calibration medium and sequence. Confirm local pressure and temperature inputs, wait for stability, and perform a separate verification. A zero procedure should be used only when the model instructions require it.
- During measurement, provide the specified sample movement and keep bubbles and deposits off the membrane. Record the probe/controller identities, settings, maintenance, reference conditions and verification result.
The exact polarization time, flow requirement, calibration points, compensation functions and acceptance limits must come from the manual supplied with the ordered configuration. For the technology comparison and record checklist, see the dissolved oxygen sensor guide.
| Brand | Apure |
|---|---|
| Model | KOG-202S |
| Measurement | Dissolved oxygen in water |
| Sensor technology | Polarographic electrochemical sensing |
| System compatibility | Use the matched probe and controller; confirm the electrical interface before ordering. |
| Order-specific specifications | Request the current KOG-202S datasheet for DO range and units, accuracy basis, process temperature and pressure limits, cable, mounting and output. |