CLI-406 Chloride Ion Selective Electrode

The Apure CLI-406 is a solid-membrane chloride ion-selective electrode system for chloride ions (Cl−) in aqueous samples. Chloride is different from free or total disinfectant chlorine; select the measuring range and interface for the actual sample.

  • Easy to operate
  • Signal response fast
  • Stable performance

At the chloride ion electrode, the internal reference liquid exudes very slowly from the microporous salt bridge at a pressure of at least 100 kpa (1Bar). Such a reference system is very stable, and the electrode life is longer than that of ordinary industrial electrodes. Easy to install: 3/4 NPT thread, easy to insert or install in pipes and tanks.

Measure chloride, rather than disinfectant chlorine

  • An ion-selective electrode develops a potential related to chloride-ion activity. Concentration reporting relies on calibration and suitable sample conditions; it does not measure free chlorine, total chlorine or ORP.
  • Use the supplied chloride standards, conditioning and calibration sequence. Keep standards and samples at comparable temperatures and use the documented ionic-strength procedure where required; do not add reagents to an online process without an approved sampling procedure.
  • Check the model’s pH window and interfering-ion limits for the water matrix. Temperature compensation does not remove interference or all activity-to-concentration effects.
  • Confirm the selected range and accuracy basis, then verify with an independent appropriate chloride reference measurement. Record standards, sample conditions, calibration and result.
  • For the listed digital/analog options, confirm the supplied wiring, power, Modbus register map and output scaling before connection.

The existing reference-system pressure and thread information above must be checked against the ordered drawing. Reference-electrolyte pressure is not a substitute for the allowable process-pressure rating in the specification table.

BrandApure
ModelCLI-406
Measuring range0-3500mg/L, 0-35000mg/L
Automatic temperature compensationAutomatic Temperature Compensation (Pt1000)
Accuracy±5%, ±0.3℃
Medium pressure≤0.2MPa
Output signalRS485 Modbus RTU & 4-20mA
Power supply12-24VDC power