A10 Digital Residual Chlorine Meter
The Apure A10 Digital Residual Chlorine Meter is an online chlorine controller used with a compatible chlorine probe. Select the supplied measurement configuration, sample-flow arrangement and control-system interface for the application.
- High precision
- High stability
- Automation
The A10 controller receives the signal from a matched chlorine probe. The membrane-covered electrochemical sensing described for the listed system belongs to the probe; the controller provides the configured display, calibration and output functions.
Commission and verify the chlorine measurement
- Confirm whether the supplied probe measures free or total chlorine, its reporting basis and the compatible controller. A chloride-ion electrode or an ORP reading is not a substitute for a chlorine-specific measurement.
- Use the specified flow cell, stable sample flow, pressure and temperature. Keep bubbles and deposits away from the sensing surface, and allow the documented conditioning and stabilization period.
- Check the sample pH and the supplied system’s pH-compensation method. Temperature compensation does not establish pH compensation or a separate pH measurement channel.
- Compare a fresh sample taken at the sensor sampling point with a suitable reference method for the same chlorine species, such as the applicable DPD procedure. Measure promptly; record sampling time, pH, temperature, flow and the reference result.
- Follow the model-specific zero/span and maintenance procedure. Verify after cleaning or membrane/electrolyte service where applicable; do not set a universal service interval for every water matrix.
Confirm the current probe and controller instructions before dosing-system integration. Free chlorine, combined chlorine and total chlorine require the appropriate measurement configuration and reference method.
The existing tables include combined sensor/controller references. Confirm the exact ordered configuration: a listed pH interval does not establish an independent pH input, and a compensation interval does not by itself establish the probe’s process-temperature rating. Panel dimensions, mains supply and relay functions describe controller options, not a bare probe. Do not apply the listed AC supply to a probe without its own approved wiring instructions.
| Product Name | A10 residual chlorine controller with a compatible sensor; listed 4-20mA configuration |
| Brand | APURE |
| Measuring range | Residual Chlorine: 0-20.00mg/L pH: 0-14pH Temp: 0-99℃ |
| Accuracy | ±1% FS |
| Stability | ±1%(FS)/24h |
| Temperature | 0~99℃(resolution ratio 0.1℃) |
| Medium pressure | 0~0.5MPa |
| Cable length | 5M |
| Output signal | 1 set of 4-20mA signal |
| Power supply | AC 220V±10% 50/60 HZ |
| Panel cut-out | 92X92mm(H*W) |
| Outline dimension | 96*96*132mm(H*W*D) |
Confirm the supplied controller functions
- The listed display, slope adjustment, factory reset and relay functions belong to the matched controller configuration; they are not functions of every bare probe.
- The existing option list cites NTC10K or PT1000 temperature inputs. Match the supplied temperature element and wiring; do not interchange them without the controller’s instructions.
- For a listed 4-20mA output, confirm measurement scaling, active/passive wiring, isolation and load. For optional RS485, obtain the actual supported protocol and register map; do not assume full Modbus compatibility from the interface name.
- Confirm alarm set points, hysteresis and relay fail-state for the process. Request the conditions and retained information for any power-failure protection claim >10 years.
The existing tables include combined sensor/controller references. Confirm the exact ordered configuration: a listed pH interval does not establish an independent pH input, and a compensation interval does not by itself establish the probe’s process-temperature rating. Panel dimensions, mains supply and relay functions describe controller options, not a bare probe. Do not apply the listed AC supply to a probe without its own approved wiring instructions.
| Brand | Apure |
|---|---|
| Model | A10 |
| Measuring range | 0-20.00mg/L |
| Automatic temperature compensation | 0-99℃ |
| Accuracy | ±1% FS |
| Stability | ±1%(FS)/24h |
| Medium pressure | 0~0.5MPa |
| Output signal | 1 set of 4-20mA signal |
| Power supply | AC 220V±10% 50/60 HZ |
| Dimensions | 96*96*132mm |
| Cut off dimensions | 92X92mm |